UPROBE-C SPECS

UPROBE-C SPECS

PRODUCT

Device Name: Wireless Probe Type Ultrasound Scanner

Model: Uprobe-C

PHYSICAL DETAILS

Dimension: 156mm x 60mm x 20mm (6.1'' x 2.4'' x 0.8'')

Weight: 260g (0.6 lbs)

Battery Life: 3 ~ 5 hours 

Charger: wireless

Recharge Time: fully charged within 2 hours

CLINICAL APPLICATIONS

Linear:

Vascular 

MSK

Pediatrics

Breast

Thyroid

Carotid

Small Organs

Nerve

Convex:

Abdomen

Cardiac

Gynecology

Obstetrics

Urology

Kidney

Lung

IMAGE

Measure: Length, Area, Angle, Heart Rate, Obstetrics  

Format: jpg, avi, and DICOM

Frame Rate: 18 frames/second

Image Adjust: BGain, TGC, DYN, Focus, Depth, Harmonic, Denoise

PROBE

Probe Frequency

Convex head: 3.5MHz/5MHz

Linear head: 7.5MHz/10MHz

Probe Element: 128

Max Depth305mm

Head Width

Convex head: 45mm

Linear head: 40mm

MODES

Scanning Mode: Electronic array

Display Mode: B, B/M; 

color doppler version: B+Color, B+PDI, B+PW

COMPATIBILITY

iOS 9.0 or later. 

iphone: 4S, 5, 5C, 5S, 6, 6 Plus, 7, 7 Plus, 8, 8 Plus, X, XR, XS, XS Max, SE (2nd gen), 11, 11 Pro, 11 Pro Max, 12 Mini, 12, 12 Pro, 12 Pro Max

ipad: Air 2, Air (3rd gen), Air (4th gen), ipad Pro series,  ipad mini (5th gen), ipad (8th gen)


Android 10 or later. 

Latest models of Android devices may not be supported.

CLEANING

Cautions

1. The probe cannot be immersed in any conductive liquid, so as not to corrode the probe and sheath.

2. The depth of the probe immersed in water shall not exceed the mark of the probe immersion line, and check the probe shell for cracks to avoid damage to internal components by liquid immersion.

3. Before cleaning the probe, make sure that the probe is powered off.

Steps:

1. After each use of the probe, use a soft cloth dampened with Isopropyl alcohol (or an appropriate hospital cleaning agent) to clean the probe.

2. Disconnect the probe from the mobile device.

3. Wipe the probe for one minute until it is visibly clean.

4. Change wipes when necessary and repeat step 3 until the probe is visibly clean.

5. Visually inspect the probe in a well-lit area to ensure all surfaces are clean. If the probe is not clean, repeat the steps above.